Products Name: ZZ15810-D Medical Syringe Liquid Leakage Tester

 

Specification:

The tester adopts a 5.7-inch color touch screen to show menus: nominal capacity of syringe,side force and axial pressure for leakage testing, and duration of applying force to the plunger, and the built- in printer can print the test report. PLC controls human machine conversation and touch screen display.

1.Product Name:Medical Syringe Testing Equipment

2.Side force: 0.25N~3N; error: within±5%

3.Axial pressure: 200kpa~300kpa; error: within±5%

4.Nominal capacity of syringe: selectable from 1ml to 60ml

5.Time of testing: 30S; error: within±1s  

 

 

Products Name: ZH15810-D Medical Tester For Determining Plunger Actuation Force

Specification:

The Tester adopts a 5.7-inch color touch screen to show menus, In use of PLC controls, nominal capacity of syringe can be selected; the screen can realize real time display of the force required to initiate movement of the plunger, the mean force during return of the plunger, the maximum and minimum force during return of the plunger, and the graph of forces required to operate plunger; test results are provided automatically, and the built -in printer can print the test report.

Load Capacity: ; error: 1N~40Nerror:within ±0.3N
Test Velocity: (100±5)mm/min
Nominal capacity of syringe: selectable from 1ml to 60ml.

 

 


Products Name: ZF15810-D Medical Syringe Air Leakage Tester

Specification:

Negative Pressure test: a manometer reading of 88kpa a blow ambient  atmospheric pressure is reached; error: within ±0.8kpa; with LED digital display

Time of testing: adjustable from 1second to 10 minuntes; within LED digital display.

( Negative pressure reading displayed on the manometer shall not change ±0.8kpa for 1 minute. )

 

bigPhoto
skype Mr. Spen Yuan
Sales Manager
Tel: 86-574-87711045
Mobile: 86- 133 3689 6428
Contact to this supplier

Didn't find what you're looking for? Post Buying Lead or contact HiSupplier Customer Service Center  for help!

Related Search

Find more related products in following catalogs on Hisupplier.com

Related Products

Company Info

NINGBO WELLMEDLAB CO.,LTD [China (Mainland)]


Business Type:Manufacturer
City: Ningbo
Province/State: Zhejiang
Country/Region: China (Mainland)

spenyspeny:   skype

You May Like: